High impedance of the reference electrode in cochlear implant patients: Three case reports and surgical considerations

0

Introduction. Electrode impedance telemetry is essential for evaluating cochlear implant (CI) function. Increased impedance of the reference electrode may compromise the performance of all active electrodes and lead to deterioration of auditory outcomes. Reports focusing specifically on reference electrode dysfunction remain limited.

Cases presentation. We report three cochlear implant recipient patients who developed excessively high reference electrode impedance with impaired auditory outcomes. The first case involved a 2-year-old girl with congenital deafness, implanted with a MED-EL Pulsar device in infancy, who developed progressively increasing impedance due to burial of the reference electrode leaves within the temporal bone. The second case concerned a 15-year-old girl implanted with a MED-EL Combi-40+ device who presented with progressive hearing deterioration and elevated impedance secondary to osseous overgrowth around the electrode. The third case involved an 18-year-old male with bilateral implants who developed fluctuating hearing secondary to the mechanical separation of the reference electrode from its leaves. In all three cases, surgical exploration identified the underlying abnormality, and corrective surgery resulted in a substantial reduction of impedance values, restoration of implant function, and improvement of auditory performance without device replacement.

Conclusions. Bone remodeling around the reference electrode may occur in patients implanted during infancy and may significantly impair cochlear implant performance. Regular long-term telemetry evaluation is recommended for pediatric cochlear implant recipients. Routine long-term telemetry monitoring is valuable for early detection of these complications and may allow successful surgical correction while preserving the implanted device.

Keywords: cochlear implant, electrode impedance, telemetry, bone remodeling, cochlear implant complications.

List of abbreviations:

MED-EL – Medical Electronic Devices

Address for correspondence: Evis BENDO
ENT Department, University Hospital Center “Mother Teresa”, Tirana, Albania
Address: Dibra’s Street, No 370, Tirana, Albania
E-mail address: drbendo@gmail.com
Phone: +35-569-2087471

Share.

Comments are closed.

Bendo E, Kiratzidis T. High impedance of the reference electrode in cochlear implant patients: three case reports and surgical considerations. Arch Balk Med Union. 2026;61(3):416-421. https://doi.org/10.31688/ABMU.2026.61.3.11